Introduction
Transient Plane Source Method (TPS)is invented by a Sweden Scientist Dr. Silas Gustafsson in 1991. The draft version of an international standard for TPS is firstly issued in 2008, and it is updated in 2022.
Model A-20 Thermal Constants Analyzer is one of the most advanced TPS instruments manufactured by FOREDA, which can measure almost all types of organic/inorganic materials and some metals/metal alloys commonly used in the industries or daily life.
Properties
-
Three parameters, thermal conductivity, thermal diffusivity, and volumetric heat capacity can be obtained with one single test
-
Test is done in seconds without balancing the electrical bridge;
-
Test Data can be saved in Excel format for further applications;
-
Software can be run in any Windows computer for off-line data analysis;
-
Very large thermal conductivity range: almost all kinds of materials can be tested;
-
Applicable for all kinds of specimens: solid, liquid, gas, powders, colloid, anisotropic; or cylindrical, triangle, films, slabs, etc.;
-
Specimens Easily prepared: only one flat surface is required;
-
Absolute measurement without calibration;
-
Possibly very large temperature range: - 270 ~ 1000℃.
Specifications
Power supply:AC 220V±5%, 100W
Thermal conductivity measurement:
range: 0.001~20 W/(m·K)
repeatability: ±1.0%
accuracy: ±2.5%
Thermal diffusivity measurement:
range: 0.1~5 mm^2/s
repeatability: ±3.5%
accuracy: ±5.0%
Volumetric heat capacity measurement:
range: 0~0.05 MJ/(m^3·K)
repeatability: ±5.5%
accuracy: ±7.0%
Test time: 2 ~ 1024 s
Specimen size requirement:
Lateral: ≥ ϕ10 mm
Thickness: ≥ 2.5 mm
Size of the Electronics: 405 (length)×356(thickness)×220 mm (height)